2. Tools, Maintenance Schedules, Procedures and Testing
Setting up your Reference Meters
With certain meter types, some sub-assembly procedures will need to be undertaken to connect
the device to your Statim calibration cassette. All cassettes are supplied with the necessary
accessories to enable you to convert your meter of choice from those recommended above, to the
appropriate fitting on the cassette.
01-103087S and 01-103088S cassettes (non S calibration) with Omega HH81A reference
meter and TJ36-CASS-116G-6-SMP-M Temperature Probe.
No sub-assembly required.
To insert the temperature probe (thermocouple) into the compression fittings, loosen the clamp
nut attached to the cassette extension fitting and insert the probe into the fitting as far as it will go.
A slight resistance will be felt as the probe passes through the seal. Tighten the clamp nut until a
steam/air tight seal is achieved.
DO NOT OVERTIGHTEN
.
NOTE: The non S calibration cassette fitting is NOT compatible with the Fluke and Heise 1/8”
(3.175mm) temperature probes.
01-106366S and 01-106367S cassettes (S-class calibration) with ALL recommended
reference meters
S-class calibration cassette kits contain a number of accessories for use with various
recommended reference meters and will need to be set up to suit the appropriate meters before
use.
Contents of calibration cassette kit:
• Cassette (2000S or 5000S as appropriate) with ‘generic’ pressure and temperature fittings
attached (see Figure 7 below)
• Accessories as follows: (inside the cassette)
1 x Pressure tube
2 x Male Swagelok B-QC4-S-2PM ‘Quick Connect’ fittings.
1 x 1/8 NPT female to ¼ NPT male adaptor
1 x 1/8” (3.175mm) probe fitting.
STAT
IM
5000/5000S/5000 G4 Service Guide