Page 94
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
4. STAT
IM 5000 Calibration
7. This Service Mode is PIN protected, enter password to continue (default password is: 7919).
8. From the service menu, select Calibration.
1. Chamber thermocouple calibration
To calibrate the chamber thermocouple:
1. To start a chamber calibration cycle, select Calibration in the Calibration submenu and
calibration with start automatically
The STATIM will run a regular UNWRAPPED cycle, but continue to show calibration
information on the LCD.
(VVV) Validation thermocouple reading
(CVO) Validation thermocouple hexadecimal offset compared with chamber reading
(TTT) Chamber thermocouple in °C
(TO) Chamber thermocouple hexadecimal offset
(DDD) Difference between validation TC and chamber temperatures in °C
STAT
IM
5000/5000S/5000 G4 Service Guide
VV
V
CVO
TTT
TO
DDD