2. Tools, Maintenance Schedules, Procedures and Testing
STAT
IM
5000/5000S/5000 G4 Service Guide
Figure 7.
S-class ‘generic’ fitting configuration (2000S cassette shown)
Setting up the temperature fittings for use with the recommended temperature
reference meter(s)
The temperature fitting (right hand fitting when facing the front of the cassette) is fitted with a
compression fitting that consists of a body, clamp nut and conical compression washer. (See
Figure 8)
There are two fittings supplied with the cassette, one for a 1/8” (3.175mm) temperature probe and
one for a 1/16” (1.6mm) temperature probe.
NOTE: The 1/16” fitting is attached to the cassette when supplied and the 1/8” fitting is in the pack
of accessories supplied with the cassette.