STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
1. Identifying STAT
IM 5000 Units
Identifying Steam Generator Types
There are three different types of steam generator (boiler) that may be encountered in STAT
IM
5000/5000S/5000 G4 units. They can be identified as below (Figure 2):
Aluminum steam generator with internal thermocouple (1995 – 2004):
Stainless steel steam generator (2004 – 2005):
Typical features:
•
70 p.s.i. pressure relief valve (PRV)
(no ring pull)
•
Thermocouple embedded in side of
steam generator
•
Software required to drive this steam
generator = R1xx/R2xx (revision
2.x/5.x PCB)
Typical features:
•
Software required to drive this steam
generator = R4xx (revision 6 PCB only)
STAT