CTS55
List of Commands/Annex C
1094.0006.55 C.17
E-8
Ramping Mode
Syntax:
CONFigure:RFGen:RAMPing:STATe <value>
Value range:
ON
Signal with power ramping
OFF
Permanent
signal
Default:
OFF
State:
RFG
Option:
B7
With Query
Output level
Syntax:
CONFigure:RFGen:POWer <numeric_value>
Value range:
(-110.0 - ext. att.) ... (-50.0 - ext. att.) dBm
Default:
-50.0 dBm
State:
RFG
Option:
B7
With Query
Output Level RFOUT2
Syntax:
CONFigure2:RFGen:POWer <numeric value>
min.
Value range:
(-80.0 - ext. Att.) ... (-20.0 - ext. Att.) dBm
Default:
depending on
level at
RF IN/OUT
State:
RFG
Option:
B7
With Query
RF Generator Level
Syntax:
CONFigure:RFGen:LEVel <value>
Value range:
ON
Switched
on
OFF
Switched
off
Default:
ON
State:
RFG
Option:
B7
With Query
Содержание 1094.0006.50
Страница 4: ......
Страница 12: ......
Страница 22: ......
Страница 24: ......
Страница 26: ......
Страница 28: ......
Страница 30: ......
Страница 41: ...CTS55 Getting Started With Autotest 1094 0006 55 3 3 E 7...
Страница 66: ......
Страница 81: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 7 E 6...
Страница 83: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 9 E 6...
Страница 106: ......
Страница 108: ......
Страница 196: ......
Страница 246: ......
Страница 250: ......
Страница 253: ...CTS55 Index 1094 0006 55 I 3 E 8 V Viewing angle 2 3 Voltage selector 2 3 Voltage selector switch 2 2...
Страница 254: ......
Страница 256: ......
Страница 343: ...CTS K6 Windows Application CTSgo 1079 2001 01 87 E 3 Fig CTSgo help window...