Getting Started With Autotest
CTS55
1094.0006.55 3.14
E-7
Step 18
Autotest Result Menu
After the test sequence has
finished the menu will have new
buttons as shown on the
opposite side. The "Restart" key
will repeat the test sequence.
The "Start Menu" key will return
to the Start Menu of the
Autotest. The "Meas" key will go
to the Power ramp Menu where
the power ramp is shown
graphically. The Phase Freq
graph will be available for
display from within the Power
ramp Menu. With the "Values"
key the display can be changed
to show the results as either
values, the tolerances or the
results as pass/fail values. The
text on the key will change to
show the next screen in the
toggle sequence. "Print Report"
prints the measurement results
into the printer protocol.