CTS55
Description of the Menus
1094.0006.55 9.27
E-7
9.8
Measurement Configuration Menus
9.8.1
Config Power Ramp
In this menu the tolerances for the power ramp measurement are defined. These are largely identical
with the relative level values for the power ramp template. The power ramp template is defined with
reference to the power used by the mobile to transmit.
The template is shown at the top of the screen. The various ranges of the template are identified with
the letters A to F. The predefined values for these ranges are shown below the template.
For range A an absolute minimum control level can also be defined.
Apart from the template the limits for the absolute power control levels of the mobile are also defined.
They consist of two values, one for the highest power control level and one for all other power control
levels.
Default
Press this key to reset all values to factory default values.
Содержание 1094.0006.50
Страница 4: ......
Страница 12: ......
Страница 22: ......
Страница 24: ......
Страница 26: ......
Страница 28: ......
Страница 30: ......
Страница 41: ...CTS55 Getting Started With Autotest 1094 0006 55 3 3 E 7...
Страница 66: ......
Страница 81: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 7 E 6...
Страница 83: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 9 E 6...
Страница 106: ......
Страница 108: ......
Страница 196: ......
Страница 246: ......
Страница 250: ......
Страница 253: ...CTS55 Index 1094 0006 55 I 3 E 8 V Viewing angle 2 3 Voltage selector 2 3 Voltage selector switch 2 2...
Страница 254: ......
Страница 256: ......
Страница 343: ...CTS K6 Windows Application CTSgo 1079 2001 01 87 E 3 Fig CTSgo help window...