CTS55
List of Commands/Annex C
1094.0006.55 C.25
E-8
Sensitive Power Measurement Offset
Syntax:
CONFigure:MODuletest:SPM:OFFSet[?]
Value range:
-
10 dB
+10 dB
Default:
2 dB
State:
MOD
Option:
Note
: This value can also be adjusted
automatically by the SPM calibration
routine.
Calibration Routine Start
Syntax:
PROCedure:CALibrate:SPM?
Value range:
"CALIBRATION_OK",
"CALIBRATION_FAILED_WRONG_FREQUENCY",
"CALIBRATION_FAILED_INVALID_SIGNAL",
"CALIBRATION_FAILED_UNSTABLE_SIGNAL"
Default:
State:
MOD
Option:
Only query
Note
: This command starts the
SPM calibration routine.