List of Commands/Annex C
CTS55
1094.0006.55 C.22
E-8
Tolerances for power time template
The level designations used in the power time template for the command
CALCulate:LIMit:POWer[:TEMPlate][:DATA] can be obtained from the following illustration:
A
B
C
D
E
F
G
H
A
Syntax:
CALCulate:LIMit:POWer[:TEMPlate][:DATA]
<numeric_value>,
<numeric_value>, <numeric_value>, <numeric_value>, <numeric_value>,
<numeric_value>, <numeric_value>, <numeric_value>, <numeric_value>
Value range:
-100.0 ... 0.0 dBm
Absolute level at A
-100.0 ... 0.0 dB
Relative level at A
-100.0 ... 0.0 dB
Relative level at B
-100.0 ... 0.0 dB
Relative level at C
0.0 ... +20.0 dB
Relative level at D
0.0 ... +5.0 dB
Relative level at E
-5.0 ... 0.0 dB
Relative level at F
-100.0 ... 0.0 dB
Relative level at G
-100.0 ... 0.0 dB
Relative level at H
Default:
GSM: -36.0
dBm
GSM850: -36.0
dBm
DCS1800: -47.0
dBm
DCS1900: -47.0
dBm
-45.0 dB
-30.0 dB
-6.0 dB
+4.0 dB
+1.0 dB
-1.0 dB
-6.0 dB
-30.0 dB
State:
ALL
Option:
none With Query
Query for observance of the tolerances for average power
Syntax:
CALCulate:LIMit:POWer[:TEMPlate]:TOLerance:MATChing?
Return:
(MATC | NMAT | INV)
State:
MCE
Option:
none Query only
Note:
returns the result from the last measurement, (N5), (N6)
Содержание 1094.0006.50
Страница 4: ......
Страница 12: ......
Страница 22: ......
Страница 24: ......
Страница 26: ......
Страница 28: ......
Страница 30: ......
Страница 41: ...CTS55 Getting Started With Autotest 1094 0006 55 3 3 E 7...
Страница 66: ......
Страница 81: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 7 E 6...
Страница 83: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 9 E 6...
Страница 106: ......
Страница 108: ......
Страница 196: ......
Страница 246: ......
Страница 250: ......
Страница 253: ...CTS55 Index 1094 0006 55 I 3 E 8 V Viewing angle 2 3 Voltage selector 2 3 Voltage selector switch 2 2...
Страница 254: ......
Страница 256: ......
Страница 343: ...CTS K6 Windows Application CTSgo 1079 2001 01 87 E 3 Fig CTSgo help window...