Remote Control
R&S
®
ESR
881
User Manual 1175.7068.02 ─ 12
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<NoMaxima>
Range:
1 to 200
Example:
INIT:CONT OFF
Switches to single sweep mode
INIT;*WAI
Starts measurement and synchronizes to end
CALC:MARK:TRAC 1
Sets marker 1 to trace 1
CALC:MARK:FUNC:FPE:SORT X
Sets the sort mode to increasing X values
CALC:MARK:FUNC:FPE 3;*WAI
Searches the 3 highest maxima for trace 1 and synchronizes to
end
CALC:MARK:FUNC:FPE:COUN?
Queries the number of maxima found
CALC:MARK:FUNC:FPE:Y?
Queries the level of maxima found
CALC:MARK:FUNC:FPE:X?
Queries the frequencies (span <> 0) or time (span = 0) of max-
ima found.
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE
<MaxNoPeaks>
This command defines the maximum number of peaks the marker peak list may con-
tain.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<MaxNoPeaks>
Maximum number of peaks to be determined.
*RST:
50
Example:
CALC:MARK:FUNC:FPE:LIST:SIZE 10
The marker peak list will contain a maximum of 10 peaks.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SEAR:AUTO
<State>
This command turns the marker peak search on and off.
This command is retained for compatibility with R&S FSP only. Use
MARKer<m>:FUNCtion:FPEaks:STAT
instead.
Remote Commands in Spectrum Analyzer Mode