Receiver Mode
R&S
®
ESR
160
User Manual 1175.7068.02 ─ 12
Remote command:
5.1.7.5
Final Measurement Control
The final measurement is started in the "Test Automation" menu . The R&S
ESR pro-
vides two methods to perform a final measurement: an automatic and interactive final
measurement.
At the start of a final measurement, the R&S
ESR opens a softkey menu to control the
measurement.
The measurement configuration is as defined in the scan table.
Sequence for automatic final measurements
Generally, the automatic final measurements requires no interactions. however, the
measurement can be interrupted, e.g. to change the mode or to analyze the signal in
detail.
► Press the "Hold Final Measurement" softkey.
The R&S
ESR interrupts the measurement. While the measurement is interrupted,
the receiver settings can be changed in order to examine the signal in detail.
The interruption also changes the contents of the "Measurement" menu. Several
actions are available:
●
Switch the mode of the final measurement ("Automatic" or "Interactive").
●
Resume the final measurement ("Measure").
The final measurement starts or continues at the next entry of the peak list.
●
Abort the final measurement ("Stop Final Measurement").
The collected data is lost.
After all frequencies in the peak list have been measured, the R&S
ESR opens the
"Final Peak List" with the results for the final measurement.
Sequence for interactive final measurements
For an interactive final measurement, the R&S
ESR initiates the following sequence.
1. The R&S
ESR tunes to the first frequency in the peak list (or the next frequency).
2. The R&S
ESR stops the measurement and positions a marker on the frequency.
3. If required, the frequency can be fine tuned, e.g. if the interferer has shifted. For
fine tuning, the bargraph display can be used to find the new peak value.
In addition, you have several options:
● Skip the current frequency ("Skip Frequency")
Positions the marker on the next frequency in the peak list without performing a
final measurement on the current frequency.
● Get maxhold result for the current frequency ("Get Maxhold")
Measurements and Result Displays