Remote Control
R&S
®
ESR
656
User Manual 1175.7068.02 ─ 12
Example:
CALC:MARK2:SGR:Y:MAX
Activates and positions marker 2 on the peak level of the y-axis.
CALC:MARK2:SGR:Y:MAX:ABV
Positions marker 2 on the next higher level found in the diagram
area above the current marker position.
Usage:
Event
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:BELow
This command positions the specified marker on the next peak level of the vertical
axis. The search includes only frames below the current marker position. It does not
change the horizontal position of the marker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m>
1...16
Selects the marker.
Example:
CALC:MARK2:SGR:Y:MAX
Activates and positions marker 2 on the peak level of the y-axis.
CALC:MARK2:SGR:Y:MAX:BEL
Positions marker 2 on the next higher level found in the diagram
area below the current marker position.
Usage:
Event
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:NEXT
This command positions the specified marker on the next peak level of the vertical
axis. The search includes frames above and below the current marker position. It does
not change the horizontal position of the marker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m>
1...16
Selects the marker.
Example:
CALC:MARK2:SGR:Y:MAX
Activates and positions marker 2 on the peak level of the y-axis.
CALC:MARK2:SGR:Y:MAX:NEXT
Positions marker 2 on the next higher level.
Usage:
Event
Remote Commands in Receiver Mode