Receiver Mode
R&S
®
ESR
196
User Manual 1175.7068.02 ─ 12
Final Measurement Time
Defines the measurement time for the final measurement.
The measurement time is the time during which the R&S
ESR measures the input sig-
nal and forms a measurement result weighted by the selected detector. The measure-
ment time does not include settling times of the synthesizer and the IF filter. The
R&S
ESR automatically waits until transients are over.
Interactive Mode
Turns interactive final measurements on and off.
Remote command:
Trace 1 to 6
Selects the characteristics of each trace.
In addition to the trace mode, you can select a detector for the scan and the final mea-
surement. For more information see
●
●
Chapter 5.1.5, "Final Measurement"
●
●
Note:
When you select detector type "None" for the final measurement, the R&S
ESR
ignores peaks found on the corresponding trace during the final measurement.
Remote command:
Trace mode:
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Scan detector:
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]
Final measurement detector:
[SENSe:]DETector<t>:FMEasurement
5.3.7.6
LISN Settings
The "LISN Settings" tab in the "Test Automation" dialog box contains functionality to
control line impedance networks.
Common Measurement Settings