Remote Control
R&S
®
ESR
767
User Manual 1175.7068.02 ─ 12
Example:
ESP:RANG3:INP:GAIN:STATe ON
Switches the preamplifier for range 3 on or off.
Manual operation:
See
[SENSe:]ESPectrum:RANGe<range>:INSert
<Mode>
This command inserts a new range before or after the specified range. The range num-
bers are updated accordingly.
Suffix:
<range>
.
1...20
range
Parameters:
<Mode>
AFTer | BEFore
Example:
ESP:RANG3:INS BEF
Inserts a new range before range 3.
ESP:RANG1:INS AFT
Inserts a new range after range 1.
Manual operation:
See
See
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:ABSolute:STARt
<Level>
This command sets an absolute limit value at the start frequency of the specified
range. Different from manual operation, this setting is independently of the defined limit
check type.
Suffix:
<range>
.
1...20
range
Parameters:
<Level>
-400 to in 400 dBm
*RST:
13 dBm
Example:
ESP:RANG1:LIM:ABS:STAR 10
Sets an absolute limit of 10 dBm at the start frequency of the
range.
Manual operation:
See
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:ABSolute:STOP
<Level>
This command sets an absolute limit value at the stop frequency of the specified
range. Different from manual operation, this setting is independent of the defined limit
check type.
Remote Commands in Spectrum Analyzer Mode