Remote Control
R&S
®
ESR
585
User Manual 1175.7068.02 ─ 12
Table 11-14: Meaning of the bits used in the STATus:QUEStionable:TRANsducer register
Bit No.
Meaning
0
Range 1
This bit is set if subrange 1 has been reached.
1
Range 2
This bit is set if subrange 2 has been reached.
2
Range 3
This bit is set if subrange 3 has been reached.
3
Range 4
This bit is set if subrange 4 has been reached.
4
Range 5
This bit is set if subrange 5 has been reached.
5
Range 6
This bit is set if subrange 6 has been reached.
6
Range 7
This bit is set if subrange 7 has been reached.
7
Range 8
This bit is set if subrange 8 has been reached.
8
Range 9
This bit is set if subrange 9 has been reached.
9
Range 10
This bit is set if subrange 10 has been reached.
10 to 13
Unused
14
Subrange limit
This bit is set when the transducer is at the point of changeover from one subrange to another.
15
This bit is always 0.
Status Registers in Spectrum Mode
This chapter contains the description of the registers and bits specific to spectrum
mode. For a comprehensive description of the other status registers see
Remote Control - Basics