Remote Control
R&S
®
ESR
842
User Manual 1175.7068.02 ─ 12
Manual operation:
See
[SENSe:]SWEep:EGATe:LENGth
<TimeInterval>
This command defines a gate length.
Parameters:
<TimeInterval>
125 ns to 30 s
*RST:
400μs
Example:
SWE:EGAT:LENG 10ms
Manual operation:
See
"Gate Length (Gate Mode Edge)"
[SENSe:]SWEep:EGATe:POLarity
<Polarity>
This command determines the polarity of the external gate signal. The setting applies
both to the edge of an edge-triggered signal and the level of a level-triggered signal.
Parameters:
<Polarity>
POSitive | NEGative
*RST:
POSitive
Example:
SWE:EGAT:POL POS
Manual operation:
See
See
[SENSe:]SWEep:EGATe:SOURce
<Source>
This command selects the signal source for gated measurements.
If an IF power signal is used, the gate is opened as soon as a signal at > -20 dBm is
detected within the IF path bandwidth (10 MHz).
For details see the
Parameters:
<Source>
EXTernal | IFPower | VIDeo | RFPower | PSEN
*RST:
IFPower
Example:
SWE:EGAT:SOUR IFP
Switches the gate source to IF power.
Manual operation:
See
See
See
See
See
See
See
Remote Commands in Spectrum Analyzer Mode