Spectrum Measurements
R&S
®
ESR
284
User Manual 1175.7068.02 ─ 12
Spurious Emissions measurements are performed using the "Spurious Emissions"
softkey in the "Measurement" menu (see
Most parameters are defined in the "Sweep List" dialog box (see
on page 289). For information on other parameters, see the corresponding soft-
Conditions for ranges
The following rules apply to ranges:
●
The minimum span of a range is 20 Hz.
●
The individual ranges must not overlap (but need not directly follow one another).
●
The maximum number of ranges is 20.
●
The maximum number of sweep points in all ranges is limited to 100001.
In order to change the start/stop frequency of the first/last range, select the appropriate
span with the [SPAN] key. If you set a span that is smaller than the overall span of the
ranges, the measurement includes only the ranges that lie within the defined span and
have a minimum span of 20 Hz. The first and last ranges are adapted to the given
span as long as the minimum span of 20 Hz is not violated.
Measurements