Receiver Mode
R&S
®
ESR
159
User Manual 1175.7068.02 ─ 12
........................................................................................ 159
Hold Scan
Interrupts the scan and opens a submenu that contains functionality to control the
scan.
Data that has already been collected is kept in the memory.
Remote command:
Continue at Rec Frequency
Resumes the scan at a set receiver frequency.
After "Hold", the receiver frequency can be set to a frequency that was already mea-
sured, i.e. which is lower than the frequency at which the scan was interrupted. As
soon as the "Continue at Rec Frequency" softkey is pressed, the scan restarts at this
frequency. This function can be used repeat part of the measurement.
If a frequency is set that is higher than the frequency at which the scan was interrupted
the scan continues at the frequency at which it was interrupted.
The scan is continued with the settings in the scan table.
Remote command:
Continue at Hold
Resumes the scan at the frequency it was interrupted.
The scan is continued with the settings in the scan table.
Remote command:
Continue at Frame
Resumes the scan at a previously recorded spectrogram frame.
The data recorded between the selected frame and the current frame is recorded
again, including minimum and maximum hold information.
Available for scans that are combined with a spectrogram measurement and the mini-
mum or maximum hold trace mode.
Remote command:
not supported
Stop Scan
Aborts the scan.
Data that has already been collected is lost.
Measurements and Result Displays