Quick Start
R&S
®
ESR
69
User Manual 1175.7068.02 ─ 12
The diagram footer can be removed from the display temporarily, see
3.4.1.5
Instrument and Status Information
Global instrument settings, the instrument status and any irregularities are indicated in
the status bar beneath the diagram.
Hiding the status bar
You can hide the status bar display, e.g. in order to enlarge the display area for the
measurement results.
1. Press the [DISPLAY] key.
2. In the "Display Settings" dialog box, select "Status Bar State: Off".
The status bar is no longer displayed.
●
To display the status bar again, select "Status Bar State: On".
SCPI command:
DISP:SBAR:STAT OFF
The following information is displayed:
Instrument status
The instrument is configured for operation with an external reference.
Error information
If errors or irregularities are detected, a keyword and an error message, if available,
are displayed in the status bar.
The following keywords are used:
UNCAL
One of the following conditions applies:
●
Correction data has been switched off.
●
No correction values are available. This occurs, for example, if a firmware
update is performed followed by a cold start of the instrument.
●
Record the correction data by performing a self alignment
OVLD
Overload of the input mixer.
●
Increase the RF attenuation (for RF input).
●
Reduce the input level (for digital input)
IFOVL
Overload of the IF signal path after the input mixer.
●
Increase the reference level.
LOUNL
Error in the instrument's frequency processing hardware was detected.
Basic Operations