Remote Control
R&S
®
ESR
642
User Manual 1175.7068.02 ─ 12
11.4.1.4
Final Measurements and Automated Test Sequence Control
............................................................................................ 642
[SENSe:]DETector<t>:FMEasurement
..............................................................................642
.......................................................................................643
........................................................................................ 643
INITiate<n>:EMITest
This command initiates an automated test sequence.
The sequence consists of a scan, a peak search and a final measurement.
Example:
INIT2:EMIT
Starts the test sequence.
Usage:
Event
INITiate<n>:FMEasurement
This command initiates a final measurement based on the peak list.
Example:
INIT2:FME
Starts the final measurement.
Usage:
Event
[SENSe:]DETector<t>:FMEasurement
<Detector>
This command selects the detector for the final measurement.
Parameters:
<Detector>
AVERage
Selects the Average detector.
CAVerage
Selects the CISPR Average detector.
CRMS
Selects the CISPR RMS detector.
NEGative
Selects the Min Peak detector.
NONE
Ignores the peaks found on that trace during the final measure-
ment.
POSitive
Selects the Max Peak detector.
QPEak
Selects the Quasipeak detector.
RMS
Selects the RMS detector.
Remote Commands in Receiver Mode