Spectrum Measurements
R&S
®
ESR
265
User Manual 1175.7068.02 ─ 12
Decim Sep ← Save Evaluation List ← List Evaluation ← Spectrum Emission
Mask
Selects the decimal separator with floating-point numerals for the ASCII Trace export
to support evaluation programs (e.g. MS-Excel) in different languages. The values '.'
(decimal point) and ',' (comma) can be set.
Remote command:
Edit Reference Range ← Spectrum Emission Mask
Opens the "Reference Range" dialog box to edit the additional settings used for SEM
measurements.
Two different power reference types are supported:
●
"Peak Power"
Measures the highest peak within the reference range.
●
"Channel Power"
Measures the channel power within the reference range (integral bandwidth
method).
If the "Channel Power" reference power type is activated, the dialog box is exten-
ded to define additional settings:
●
"Tx Bandwidth"
Defines the bandwidth used for measuring the channel power:
minimum span ≤ value ≤ span of reference range
●
"RRC Filter State"
Activates or deactivates the use of an RRC filter.
●
"RRC Filter Settings"
Sets the alpha value of the RRC filter. This window is only available if the RRC filter
is activated.
For further details refer to
Remote command:
[SENSe:]ESPectrum:FILTer[:RRC][:STATe]
[SENSe:]ESPectrum:FILTer[:RRC]:ALPHa
Measurements