The output of the tracking generator is connected to the DUT via a cable equipped with
a male N connector. The female connector is available only with the tracking generator
option (R&S
FSV-B9).
Risk of damaging the tracking generator
To avoid damaging the tracking generator, make sure that the reverse power (i.e. input
power caused, for example, by reflections in the test setup) never exceeds 1
W RF
power. Do not apply a DC voltage at the output.
For details, refer to the data sheet.
Sensitive DUTs concerning matching
For DUTs with sensitive RF characteristics with regard to matching (VSWR) at the
input, insert a 10
dB attenuator between the DUT and the tracking generator.