Remote Control
R&S
®
ESR
809
User Manual 1175.7068.02 ─ 12
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:AVERage:RESult?
This command queries the average standard deviation of the time domain power. The
query is only possible if averaging has been activated previously using
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage
.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Return values:
<StandardDeviation> Standard deviation of the signal during the measurement time.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:SDEV ON
Switches on the function.
CALC:MARK:FUNC:SUMM:AVER ON
Switches on the calculation of average.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:MEAN:SDEV:RES?
Outputs the result.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:PHOLd:RESult?
This command queries the maximum standard deviation of the time domain power.
The query is only possible if the peak hold function has been activated previously using
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Return values:
<StandardDeviation> Standard deviation of the signal during the measurement time.
Remote Commands in Spectrum Analyzer Mode