Remote Control
R&S
®
ESR
709
User Manual 1175.7068.02 ─ 12
CALCulate<n>:LIMit<k>:CONTrol[:DATA]
<LimitLinePoints>
This command defines the horizontal definition points of a lower limit line.
Suffix:
<n>
.
Selects the measurement window.
<k>
1...8
Selects the limit line.
Parameters:
<LimitLinePoints>
Variable number of x-axis values.
Note that the number of horizontal values has to be the same as
the number of vertical values set with
ESR either adds missing val-
ues or ignores surplus values.
The unit is Hz or s, depending on
*RST:
- (CALC:LIM is set to OFF)
Example:
CALC:LIM2:CONT 1 MHz,30 MHz,100 MHz,300 MHz,1
GHz
Defines 5 reference values for the x-axis of limit line 2.
CALC:LIM2:CONT?
Outputs the reference values for the x-axis of limit line 2 separa-
ted by a comma.
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol:DOMain
<SpanSetting>
This command selects the domain of the limit line.
Suffix:
<n>
.
Selects the measurement window.
<k>
1...8
Selects the limit line.
Parameters:
<SpanSetting>
FREQuency | TIME
*RST:
FREQuency
Example:
CALC:LIM2:CONT:DOM TIME
Defines zero span for the x-axis of limit line 2.
CALCulate<n>:LIMit<k>:CONTrol:MODE
<Mode>
This command selects the horizontal limit line scaling.
Suffix:
<n>
.
Selects the measurement window.
Remote Commands in Receiver Mode