Example:
CALC:LIM2:CONT:SHIF 50KHZ
Shifts all reference values of limit line 2 by 50 kHz.
CALCulate<n>:LIMit<k>:CONTrol:SPACing
<InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from frequency points.
Suffix:
<n>
.
Selects the measurement window.
<k>
1...8
Selects the limit line.
Parameters:
<InterpolMode>
LINear | LOGarithmic
*RST:
LIN
Example:
CALC:LIM:CONT:SPAC LIN
CALCulate<n>:LIMit<k>:LOWer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
If the measured values are smaller than the LOWer limit line, the limit check is violated.
Suffix:
<n>
.
irrelevant
<k>
1...8
Selects the limit line.
Parameters:
<LimitLinePoints>
Variable number of x-axis values.
Note that the number of horizontal values has to be the same as
the number of vertical values set with
ESR either adds
missing values or ignores surplus values.
The unit depends on
*RST:
(LIMit:STATe is set to OFF)
Example:
CALC:LIM2:LOW -30,-40,-10,-40,-30
Defines 5 lower limit values for limit line 2 in the preset unit.
CALC:LIM2:LOW?
Outputs the lower limit values of limit line 2 separated by a
comma.
Manual operation:
See
CALCulate<n>:LIMit<k>:LOWer:MARGin
<Margin>
This command defines an area around a lower limit line where limit check violations
are still tolerated.
Remote Commands in Receiver Mode