Remote Control
R&S
®
ESR
846
User Manual 1175.7068.02 ─ 12
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
<Position>
This command defines the position of the reference level on the display grid..
Suffix:
<n>
.
Selects the measurement window.
<t>
irrelevant
Parameters:
<Position>
0 PCT corresponds to the lower display border, 100% corre-
sponds to the upper display border.
Range:
0 to 100
*RST:
100 PCT = frequency domain, 50 PCT = time
domain
Default unit: PCT
Example:
DISP:TRAC:Y:RPOS 50PCT
Manual operation:
See
See
INPut:ATTenuation:AUTO
<State>
This command automatically couples the input attenuation to the reference level (state
ON) or switches the input attenuation to manual entry (state OFF).
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
INP:ATT:AUTO ON
Couples the attenuation set on the attenuator to the reference
level.
Manual operation:
See
UNIT<n>:POWer
<Unit>
This command selects the default unit.
The unit is the same for al measurementl windows.
Parameters:
<Unit>
DBM | V | A | W | DBPW | WATT | DBPT | DBUV | DBMV |
VOLT | DBUA | AMPere
*RST:
DBM
Example:
UNIT:POW DBUV
Sets the power unit to dBm.
Usage:
SCPI confirmed
Remote Commands in Spectrum Analyzer Mode