Remote Control
R&S
®
ESR
936
User Manual 1175.7068.02 ─ 12
Return values:
<CalibrationData>
String containing the calibration data.
Example:
CAL:RES?
would return, e.g.
Total Calibration Status:
PASSED, Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Usage:
Query only
Manual operation:
See
CALibration:STATe
<State>
This command includes or excludes calibration data in the actual measurement.
Parameters:
<State>
ON | OFF
Example:
CAL:STAT OFF
Sets up the instrument to ignore the calibration data.
Usage:
SCPI confirmed
DIAGnostic<n>:SERVice:INPut:PULSed:CFRequency
<Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic<n>:SERVice:INPut[:SELect]
Suffix:
<n>
.
irrelevant
Parameters:
<Frequency>
Possible frequencies of the calibration signal are fix.
If you define a frequency that is not available, the R&S
ESR
uses the next available frequency. Example: a frequency of
20
MHz is rounded up to the next available frequency
(21.33
MHz).
*RST:
64 MHz
Example:
DIAG:SERV:INP:PULS:CFR 128 MHz
Defines a calibration signal frequency of 128 MHz.
Usage:
SCPI confirmed
Manual operation:
See
System Configuration