Remote Control
R&S
®
ESR
830
User Manual 1175.7068.02 ─ 12
Manual operation:
See
[SENSe:]ADJust:LEVel
This command automatically sets the optimal reference level for the current measure-
ment.
You can define a threshold that the signal must exceed before the reference level is
adjusted, see
[SENSe:]ADJust:CONFiguration:HYSTeresis:UPPer
[SENSe:]ADJust:CONFiguration:HYSTeresis:LOWer
.
Example:
ADJ:LEV
Manual operation:
See
See
See
11.5.2.3
Bandwidth Settings
The following commands select the measurement filter bandwidths and characteristics.
Selecting the Resolution Bandwidth
Commands useful to configure the RBW described elsewhere:
●
[SENSe:]BANDwidth|BWIDth[:RESolution]
●
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE
[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO
..............................................................830
[SENSe:]BANDwidth|BWIDth[:RESolution]:FFT
.................................................................831
[SENSe:]BANDwidth|BWIDth[:RESolution]:RATio
..............................................................831
[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO
<State>
This command couples and decouples the resolution bandwidth to the span.
The automatic coupling adapts the resolution bandwidth to the current frequency span
according to the relationship between frequency span and resolution bandwidth.
Use
[SENSe:]BANDwidth|BWIDth[:RESolution]:RATio
to define the ratio
RBW/span.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
BAND:AUTO OFF
Switches off the coupling of the resolution bandwidth to the
span.
Manual operation:
See
See
See
Remote Commands in Spectrum Analyzer Mode