CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:LIMit<k>:DELTa?
This command queries the vertical distance from the deltamarker position to the limit
line. The unit is dB.
If the deltamarker has been assigned to a different trace than the limit line, the com-
mand returns -200.
Suffix:
<n>
.
irrelevant
<m>
1...16
deltamarker
<k>
1...8
limit line
Return values:
<Amplitude>
Vertical distance to the limit line in dB.
Example:
CALC:DELT3:FUNC:FME:LIM2:DELT?
Queries the distance of deltamarker 3 to the second limit line.
Usage:
Query only
Manual operation:
See
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:RESult?
<Result>
This command queries the result of the EMI measurement at the deltamarker position.
Suffix:
<n>
.
irrelevant
<m>
1...16
marker
Return values:
<Result>
Power level in dB related to the reference marker you have set
for that delta marker.
Example:
CALC:DELT3:FUNC:FME:RES?
Queries the result of delta marker 3.
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:LIMit<k>:CONDition ?
<Condition>
This command queries the condition of a marker position in relation to a certain limit
line.
Suffix:
<n>
.
irrelevant