Remote Control
R&S
®
ESR
812
User Manual 1175.7068.02 ─ 12
●
Setting up the instrument and querying the result list after synchronization to the
end of measurement: With this method the control computer can be used for other
activities while the instrument is performing the measurement at the cost of addi-
tional time needed for synchronization via service request.
...............................................................................................812
................................................................................... 812
...................................................................................... 813
....................................................................................... 813
[SENSe:]MPOWer:FTYPe
<FilterType>
This command selects the filter type for pulse power measurements.
Parameters:
<FilterType>
NORMal | CFILter | RRC
Example:
SENSe:MPOWer:FTYPe CFILter
[SENSe:]MPOWer:RESult[:LIST]?
This command queries the results of the pulse power measurement.
This command may be used to obtain measurement results in an asynchronous way,
using the service request mechanism for synchronization to the end of the measure-
ment.
If there are no results, the command returns an error.
Parameters:
<PulsePower>
List of pulse powers.
The number of values depends on the number of pulses you
have been measuring.
The unit is dBm.
Example:
*ESE 1
*SRE 32
Configuration of status reporting systems for the generation of
an SRQ on operation complete
SENSe:MPOWer 935.2MHz,1MHz,434us,VIDEO,50PCT,
5us,MEAN,20;*OPC
Configuring and starting the measurement
…
Further actions of the control computer during measurement
On SRQ:
Response to service request
SENSe:MPOWer:RESult?
Usage:
Query only
Remote Commands in Spectrum Analyzer Mode