Remote Control
R&S
®
ESR
874
User Manual 1175.7068.02 ─ 12
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:DELT:FUNC:PNO ON
Switches on the phase-noise measurement with all delta mark-
ers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm
Manual operation:
See
Marker Demodulation
Commands useful to configure the marker demodulation described elsewhere:
●
●
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
....................................874
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff
..........................................874
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect
............................................875
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe]
........................................... 875
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
<State>
This command turns continuous demodulation of the signal at the marker position in
the frequency domain on and off.
Thus acoustic monitoring of the signals can be performed.
In the time domain continuous demodulation is always on.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC2:MARK3:FUNC:DEM:CONT ON
Switches on the continuous ' demodulation.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff
<Duration>
This command defines for how long the the signal at the marker position is demodula-
ted.
Remote Commands in Spectrum Analyzer Mode