I/Q Analyzer
R&S
®
ESR
466
User Manual 1175.7068.02 ─ 12
Range Log 5 dB ← Range
Sets the level display range to 5 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISP:WIND:TRAC:Y 5DB
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Range Log 1 dB ← Range
Sets the level display range to 1 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISP:WIND:TRAC:Y 1DB
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Range Log Manual ← Range
Opens an edit dialog box to define the display range of a logarithmic level axis man-
ually.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Range Linear % ← Range
Selects linear scaling for the level axis in %.
The grid is divided into decadal sections.
Markers are displayed in the selected unit ("Unit" softkey). Delta markers are displayed
in % referenced to the voltage value at the position of marker 1. This is the default set-
ting for linear scaling.
Remote command:
DISP:TRAC:Y:SPAC LIN
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing
Range Lin. Unit ← Range
Selects linear scaling in dB for the level display range, i.e. the horizontal lines are
labeled in dB.
Markers are displayed in the selected unit ("Unit" softkey). Delta markers are displayed
in dB referenced to the power value at the position of marker 1.
Softkeys of the Amplitude Menu in I/Q Analyzer Mode