Remote Control
R&S
®
ESR
668
User Manual 1175.7068.02 ─ 12
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing
<ScalingType>
This command selects the scaling of the y-axis.
Suffix:
<n>
.
Selects the measurement window.
<t>
irrelevant
Parameters:
<ScalingType>
LOGarithmic
Logarithmic scaling.
LINear
Linear scaling in %.
LDB
Linear scaling in dB.
*RST:
LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Select a linear scale.
Manual operation:
See
See
See
See
See
See
See
See
INPut:ATTenuation
<Attenuation>
This command defines the attenuation level at the RF input.
To protect the input mixer, attenuation levels of 10
dB or less are possible only if you
have turned the input protection off with
INPut:ATTenuation:PROTection[:
.
Parameters:
<Attenuation>
Range:
0 dB to 75 dB
Increment: 5 dB
*RST:
10 dB
Example:
INP:ATT 40dB
Defines an attenuation level of 40 dB.
Manual operation:
See
INPut:ATTenuation:AUTO
<State>
This command turns automatic configuration of the attenuation on and off.
Remote Commands in Receiver Mode