1xEV-DO Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
61
User Manual 1173.9511.02 ─ 11
Refer to
chapter 2.2.8, "Inter-RAT Handover"
on page 25 for detailed information
about inter-RAT handover.
Remote command:
2.4.3 Using the Shortcut Softkeys
When using the 1xEV-DO signaling application and a 1xEV-DO measurement in paral-
lel, it is recommended to use a shortcut softkey to switch to the measurement.
Using one of these softkeys ensures that the measurement is configured compatible
with the settings of the signaling application. When you use the softkeys to switch to
the "TX Measurements", the combined signal path scenario is activated automatically
in the measurement.
Consequences:
●
The measurement and the signaling application can be used in parallel, i.e. both
forward signal transmission and measurement can be switched on.
●
The signaling RF settings are also used for the measurement.
●
The AT signal info settings of the measurement are configured compatible with the
signaling application.
●
The frame trigger signal provided by the signaling application is selected as trigger
source by the measurement.
●
Additional softkeys and hotkeys are displayed in the measurement, so that the sig-
naling application can be controlled and configured from the measurement.
If the softkey label equals "Go to...", the softkey opens a dialog box with a list of all
available 1xEV-DO measurements. If the softkey label indicates a measurement name,
this measurement has been assigned to the softkey as fixed target, see
.
Three shortcut softkeys are available and can be set to different fixed targets.
GUI Reference