1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
443
User Manual 1173.9511.02 ─ 11
Manual operation:
See
CONFigure:EVDO:MEAS<i>:OLTR:PSTep
<PowerStep>
Defines the size of the power steps, i.e. the increases and decreases in the total BSS
power during the OLTR measurement.
Parameters:
<PowerStep>
The power step is relative to the measured reference power.
Range:
0 dB to 40 dB
*RST:
20 dB
Default unit: dB
Example:
See
OLTR: Specifying Measurement-Specific Settings
Firmware/Software:
V3.0.10
Manual operation:
See
CONFigure:EVDO:MEAS<i>:OLTR:PSTep:DIRection
<PStepDirection>
Defines the direction of the first power step within an OLTR measurement. For each
subsequent power step the direction is toggled.
Parameters:
<PStepDirection>
DOWN | UP
*RST:
DOWN
Default unit: dB
Example:
See
OLTR: Specifying Measurement-Specific Settings
Firmware/Software:
V3.0.10
Manual operation:
See
CONFigure:EVDO:MEAS<i>:OLTR:RPINterval
<RefPowInterval>
Gets the duration of the reference power interval, i.e. the interval that is used to calcu-
late the AT reference power for the subsequent power step.
Parameters:
<RefPowInterval>
The time as the number of slots: from 3 (= 5 ms) to 24 (=40 ms)
Range:
3 to 24
*RST:
6
Example:
See
OLTR: Specifying Measurement-Specific Settings
Firmware/Software:
V3.0.10
Manual operation:
See
Command Reference
135-1095-0799