1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
444
User Manual 1173.9511.02 ─ 11
CONFigure:EVDO:MEAS<i>:OLTR:RPINterval:TIME?
Gets the duration of the reference power interval, i.e. the interval that is used to calcu-
late the AT reference power for the subsequent power step.
Return values:
<RefPowInterval>
Range:
5 ms to 40 ms
Increment: 5 ms
Default unit: ms
Example:
See
OLTR: Specifying Measurement-Specific Settings
Usage:
Query only
Firmware/Software:
V3.0.10
Manual operation:
See
CONFigure:EVDO:MEAS<i>:OLTR:GINTerval
<GuardInterval>
Defines the duration of the guard intervals, i.e. the intervals succeeding the OLTR eval-
uation intervals and preceding the reference power intervals.
Parameters:
<GuardInterval>
The duration of the guard interval as number of power control
groups (1.25 ms).
Range:
3 to 60
Increment: 3
*RST:
18
Example:
See
OLTR: Specifying Measurement-Specific Settings
Firmware/Software:
V3.0.10
V3.0.20: range of "GuardInterval" changed
Manual operation:
See
CONFigure:EVDO:MEAS<i>:OLTR:GINTerval:TIME?
Gets the duration of the guard intervals, i.e. the intervals succeeding the OLTR evalua-
tion intervals and preceding the reference power intervals.
Return values:
<GuardInterval>
Range:
5 ms to 100 ms
Example:
See
OLTR: Specifying Measurement-Specific Settings
Usage:
Query only
Firmware/Software:
V3.0.10
V3.0.20: range of "GuardInterval" changed
Manual operation:
See
Command Reference