1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
325
User Manual 1173.9511.02 ─ 11
Return values:
<TriggerList>
Comma separated list of all supported values. Each value is rep-
resented as a string.
Example:
See
Configuring the Trigger System
Usage:
Query only
Firmware/Software:
V1.0.15.20
Manual operation:
See
TRIGger:EVDO:MEAS<i>:MEValuation:SOURce
<TriggerSource>
Selects the source of the trigger events. Some values are always available in this firm-
ware application. They are listed below. Depending on the installed options additional
values may be available. A complete list of all supported values can be displayed using
TRIGger:...:CATalog:SOURce?
.
Parameters:
<TriggerSource>
'Free Run':
Free Run (untriggered)
'IF Power':
Power trigger (received RF power)
'IF Auto Sync':
Power trigger auto syncronized
'Base1: External TRIG A':
External trigger fed in at TRIG A
connector
'Base1: External TRIG B':
External trigger fed in at TRIG B
connector
*RST:
'Free Run'
Example:
See
Configuring the Trigger System
Firmware/Software:
V1.0.15.20
Manual operation:
See
TRIGger:EVDO:MEAS<i>:MEValuation:SLOPe
<TriggerSlope>
Qualifies whether the trigger event is generated at the rising or at the falling edge of
the trigger pulse (valid for external trigger sources).
Parameters:
<TriggerSlope>
REDGe | FEDGe
REDGe:
Rising edge
FEDGe:
Falling edge
*RST:
REDG
Example:
See
Configuring the Trigger System
Firmware/Software:
V1.0.4.11
Manual operation:
See
Command Reference
深圳德标仪器