CDMA2000 Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
578
User Manual 1173.9511.02 ─ 11
4.3.13.3
Limit Settings
This section defines limits for the results of the "RX Measurement FER".
The conformance test specification 3GPP2 C.S0043-0 specifies a variety of test cases
related to the FER limits. Adjust the value in accordance with your measurement.
Fig. 4-65: Limit settings
................................................................................................578
................................................................................... 578
FER FCH / FER SCH0
Configures limits for FER FCH and FER SCH0 measurements.
Max FER ← FER FCH / FER SCH0
Sets the maximum FER allowed before indicating an error.
Remote command:
Min Confidence Level ← FER FCH / FER SCH0
Sets the minimum confidence level of error rate that must be met without indicating an
error.
Remote command:
4.3.13.4
FER FCH / FER SCH0 View (Tab)
This test data view shows measurement results and settings for Frame Error Rate
(FER) measurements on the FCH. FER measures the percentage of frame errors over
the total number of frames received. The R&S
CMW monitors the FCH and SCH0.
Views of both channels are analogical. For more details refer to
"Frame Error Rate (FCH / SCH0) Measurement"
GUI Reference