1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
287
User Manual 1173.9511.02 ─ 11
Assign Views
Selects the view types to be displayed in the "Overview" dialog. The R&S
CMW does
not evaluate the results for disabled views. Therefore, limiting the number of assigned
views can speed up the measurement.
Remote command:
etc.
Code Domain
Configures the parameters for the CDP and CDE measurements.
Long Code Mask I/Q ← Code Domain
Specify the I/Q-Long Code Masks used in the analyzed reverse link signal. For the
measurements to work both long code masks have to be set according to the analyzed
(reverse) signal from the AT under test.
Remote command:
Reference Power Mode ← Code Domain
Select the reference power for the measurement. This setting determines whether the
channel power (in dB) is measured relative to the pilot power or to the total channel
power (AT power). It also influences the vertical axis of the measurement diagrams.
Note that the pilot channel is always present on W
0
16
.
Remote command:
Adjacent Channel Power
For all active reverse link carriers, this function enables either the ACP measurement
at up to 10 negative and up to 10 positive frequency offsets, or the extended ACP
measurement at up to 20 negative and up to 20 positive frequency offsets from the
nominal carrier frequency. Each enabled frequency offset is displayed in the "Adjacent
Channel Power (ACP)" diagram in "Spectrum" view with the set resolution bandwidth
(see
chapter 3.2.8.4, "Multi Evaluation Detailed Views: Spectrum"
Absolute ACP in dBm and the ACP in dBc relative to carrier power (i.e. relative to nar-
rowband power for single carrier and to wideband power for multi-carrier measure-
ment) can be displayed. The unit of measured ACP is selectable per frequency offset
(dBc or dBm).