1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
308
User Manual 1173.9511.02 ─ 11
FETCh:EVDO:MEAS:MEValuation:TRACe:EVMagnitude:AVERage?
FETCh:EVDO:MEAS:MEValuation:MODulation:SDEViation?
FETCh:EVDO:MEAS:MEValuation:ACP:MAXimum?
FETCh:EVDO:MEAS:MEValuation:STATe:ALL?
3.4.1.7
OLTR: Specifying Measurement-Specific Settings
The 1xEV-DO OLTR measurement is programmed as follows:
●
The measurement is controlled by SCPI commands with the following syn-
tax:
...EVDO:MEAS:OLTR...
.
●
Use general commands of the type
...:EVDO:MEAS...
to define the signal rout-
ing and perform RF and analyzer settings.
●
Trigger setting commands are the same as by the multi evaluation measurement,
see
Configuring the Trigger System
●
After a
*RST
, the measurement is switched off. Use
READ:EVDO:MEAS:OLTR...?
to initiate a single-shot measurement and retrieve
the results. You can also start the measurement using
INIT:EVDO:MEAS:OLTR
and retrieve the results using
FETCh:EVDO:MEAS:OLTR...?
/
FETCh:EVDO:MEAS:RPINterval...?
.
// *************************************************************************
// Define measurement timeout, no. of sequences for
// single-shot measurement, power step 30 dB, direction up,
// reference power interval 20 ms, guard interval 50 ms,
// measure on exception and initial lower OLTR limit.
// *************************************************************************
CONFigure:EVDO:MEAS:OLTR:TOUT 5
CONFigure:EVDO:MEAS:OLTR:SEQuence 5
CONFigure:EVDO:MEAS:OLTR:PSTep 30
CONFigure:EVDO:MEAS:OLTR:PSTep:DIRection UP
CONFigure:EVDO:MEAS:OLTR:RPINterval 12
CONFigure:EVDO:MEAS:OLTR:RPINterval:TIME?
CONFigure:EVDO:MEAS:OLTR:GINTerval 30
CONFigure:EVDO:MEAS:OLTR:GINTerval:TIME?
CONFigure:EVDO:MEAS:OLTR:MOEXception ON
CONFigure:EVDO:MEAS:OLTR:LIMit:ILOWer -2
3.4.1.8
OLTR: Single-Shot Measurement
// *************************************************************************
// Start single-shot measurement and return results for the open loop time
// response power up, the limit violation and the reference power trace
// for sequence 2.
// Query the measurement state (should be "RDY").
// Query measurement state and substates (should be "RUN,ADJ,ACT").
// *************************************************************************
INITiate:EVDO:MEAS:OLTR
FETCh:EVDO:MEAS:OLTR:SEQuence2:TRACe:UP?
Programming