1xEV-DO Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
130
User Manual 1173.9511.02 ─ 11
2.5.1.13
Configuring Message Monitoring
// *****************************************************************************
// Enable message monitoring for 1xEV-DO. Select address number 2 from the global
// logging PC address pool. Query the corresponding IP address string.
// *****************************************************************************
CONFigure:EVDO:SIGN:MMONitor:ENABle ON
CONFigure:EVDO:SIGN:MMONitor:IPADdress IP2
CONFigure:EVDO:SIGN:MMONitor:IPADdress?
2.5.2 RX Measurements
The "PER" and "Throughput" RX measurements provided by the 1xEV-DO signaling
application are programmed as follows:
●
After a
*RST
, the measurement is switched off.
●
Select the appropriate test application and set up a connection to the AT using the
1xEV-DO signaling application.
●
Selectively enable or disable the measurements (see
●
Start the PER measurement using
INIT:EVDO:SIGN:PER
and retrieve the results
using
FETCh:EVDO:SIGN:RXQuality:FLPer?
and
FETCh:EVDO:SIGN:RXQuality:RLPer?
.
●
Start the throughput measurement using
INIT:EVDO:SIGN:THRoughput
and
retrieve the results using
FETCh:EVDO:SIGN:RXQuality:FLPFormance?
and
FETCh:EVDO:SIGN:RXQuality:RLPFormance?
.
The examples in this section focus on commands directly related to the RX Measure-
ment. For general configuration of the signaling application refer to
2.5.2.1
General Measurement Settings
// *************************************************************************
// System-Reset
// *************************************************************************
*RST; *OPC?
*CLS; *OPC?
// *************************************************************************
// Set the update period for all RX Measurement evaluations to 1s,
// switch on a forward link PER view, reverse link throughput
// view and a data view.
// *************************************************************************
CONFigure:EVDO:SIGN:RXQuality:UPERiod 1
CONFigure:EVDO:SIGN:RXQuality:RESult:FLPer ON
CONFigure:EVDO:SIGN:RXQuality:RESult:RLPFormance ON
CONFigure:EVDO:SIGN:RXQuality:RESult:IPSTatistics ON
Programming