1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
318
User Manual 1173.9511.02 ─ 11
ROUTe:EVDO:MEAS<i>:SCENario:CSPath
..................................................................... 319
ROUTe:EVDO:MEAS<i>:SCENario:CATalog:CSPath?
...................................................... 319
CONFigure:EVDO:MEAS<i>:RFSettings:EATTenuation
.....................................................320
ROUTe:EVDO:MEAS<i>?
Returns the configured routing settings.
Return values:
<Scenario>
SALone | CSPath
SALone
: Standalone (Non Signaling)
CSPath
: Combined Signal Path
<Controller>
Controlling application while scenario CSPath is active.
<RXConnector>
RF1C | RF2C | RF3C | RF4C
RF 1 COM to RF 4 COM front panel connectors.
<RXConverter>
RX1 | RX2 | RX3 | RX4
RX module for the input path.
Usage:
Query only
Firmware/Software:
V2.0.10
Manual operation:
See
ROUTe:EVDO:MEAS<i>:SCENario?
Returns the active scenario.
Return values:
<Scenario>
SALone | CSPath
SALone
: Standalone (Non Signaling)
CSPath
: Combined Signal Path
*RST:
STAN
Usage:
Query only
Firmware/Software:
V1.0.15.20
V2.0.10: New values for parameter
<Scenario>
, parameters
<RXConnector>
and
<RFConverter>
are moved to command
ROUTe:EVDO:MEAS
.
Manual operation:
See
ROUTe:EVDO:MEAS<i>:SCENario:SALone
<RXConnector>, <RXConverter>
Activates the Standalone (SA) scenario and selects the RF input path for the measured
RF signal, i.e. the RF connector and the RX module.