Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-163
dmm.measure.limit[Y].high.value
This attribute specifies the upper limit for a limit test.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
1 for most functions; see
Details
Usage
highLimit
= dmm.measure.limit[
Y
].high.value
dmm.measure.limit[
Y
].high.value =
highLimit
highLimit
The value of the upper limit (-1e+12 to 1e+12)
Y
Limit number:
1
or
2
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This command sets the high limit for the limit
Y
test for the selected measurement function. When limit
Y
testing is enabled, the instrument generates a fail indication when the measurement value is more
than this value.
Default is 0.8 for limit 1 when the diode function is selected; 10 when the continuity function is
selected. The default for limit 2 for the diode and continuity functions is 1.
Example
See
dmm.measure.limit[Y].low.value
(on page 8-164) for an example of how to use this command.
Also see
dmm.digitize.limit[Y].high.value
(on page 8-161)
dmm.measure.limit[Y].low.value
(on page 8-164)