Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-93
dmm.digitize.limit[Y].autoclear
This attribute indicates if the test result for limit
Y
should be cleared automatically or not.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
dmm.ON
Usage
value
= dmm.digitize.limit[
Y
].autoclear
dmm.digitize.limit[
Y
].autoclear =
value
value
The auto clear setting:
•
Disable:
dmm.OFF
•
Enable:
dmm.ON
Y
Limit number:
1
or
2
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
When auto clear is set to on for a measure function, limit conditions are cleared automatically after
each measurement. If you are making a series of measurements, the instrument shows the limit test
result of the last measurement for the pass or fail indication for the limit.
If you want to know if any of a series of measurements failed the limit, set the auto clear setting to off.
When this set to off, a failed indication is not cleared automatically. It remains set until it is cleared
with the clear command.
The auto clear setting affects both the high and low limits.
Example
dmm.digitize.func = dmm.FUNC_DIGITIZE_VOLTAGE
dmm.digitize.limit[1].autoclear = dmm.OFF
Turns off autoclear for limit 1 when
measuring digitize voltage.
Also see
(on page 8-95)
dmm.measure.limit[Y].autoclear
(on page 8-159)