Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-252
DMM7510-901-01 Rev. B / May 2015
status.standard.event
This attribute returns the contents of the Standard Event Status Register set of the status model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
status.preset()
Not applicable
0
Usage
standardRegister
= status.standard.event
standardRegister
The status of the standard event status register
Details
When this command returns zero (0), no bits are set. You can send 0 to clear all bits.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register.
Bit
Decimal
value
Constant
When set, indicates the following has occurred:
0
1
status.standard.OPC
All pending selected instrument operations are complete
and the instrument is ready to accept new commands.
The bit is set in response to an
(on page 6)
command or TSP
(on page 8-230) function.
1
2
Not used
Not used.
2
4
status.standard.QYE
Attempt to read data from an empty Output Queue.
3
8
Not used
Not used.
4
16
Not used
Not used.
5
32
Not used
Not used.
6
64
Not used
Not used.
7
128
status.standard.PON
The instrument has been turned off and turned back on
since the last time this register was read.
Command errors include:
•
IEEE Std 488.2 syntax error:
The instrument received a message that does not follow the defined
syntax of the IEEE Std 488.2 standard.
•
Semantic error:
The instrument received a command that was misspelled or received an optional IEEE
Std 488.2 command that is not implemented in the instrument.
•
GET error:
The instrument received a Group Execute Trigger (GET) inside a program message.
Example
print(status.standard.event)
May return the value
129
, showing that the Standard Event
Status Register contains binary
10000001
Also see
(on page 3)
(on page 14)