[:SENSe[1]]:<function>:ATRigger:EDGE:LEVel
This command defines the signal level that generates the analog trigger event for the edge trigger mode.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0
Usage
[:SENSe[1]]:<function>:ATRigger:EDGE:LEVel <setting>
[:SENSe[1]]:<function>:ATRigger:EDGE:LEVel?
<function>
The function to which the setting applies; see
Functions
<setting>
The signal level that generates the trigger event
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command is only available when the analog trigger mode is set to edge.
The edge level can be set to any value in the active measurement range. See the Model DMM7510
specifications for more information on the resolution and accuracy of the analog trigger.
To use the analog trigger with the measure functions, a range must be set (you cannot use
autorange) and autozero must be disabled.
Example
FUNC "CURR"
CURR:RANGE 3
CURR:AZER OFF
CURR:ATR:MODE EDGE
CURR:ATR:EDGE:LEV 2.5
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to edge.
Set the analog trigger level to 2.5 A.
Also see
(on page 3-64)
[:SENSe[1]]:<function>:ATRigger:MODE
[:SENSe[1]]:<function>:AZERo[:STATe]
(on page 6-72)
[:SENSe[1]]:<function>:SENSe:RANGe[:UPPer]