Example 2
*RST
DIG:LINE5:MODE TRIG,OUT
TRIG:DIG5:OUT:STIM TIM3
TRIG:TIM3:DEL 3e-3
TRIG:TIM3:COUNT 5
TRIG:TIM3:STAT ON
Configure timer 3 to generate 5 pulses on digital
I/O line 5 that are 3 ms apart.
Example 3
*RST
DIG:LINE3:MODE TRIG,IN
DIG:LINE5:MODE TRIG,OUT
TRIG:DIG5:OUT:STIM TIM3
TRIG:TIM3:DEL 3e-3
TRIG:TIM3:COUNT 5
TRIG:TIM3:START:STIM DIG3
TRIG:TIM3:STAT ON
Configure timer 3 to generate 5 pulses on digital
I/O line 5 that are 3 ms apart when a digital input
is detected on digital line 3.
Also see
None