Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-125
[:SENSe[1]]:TRIGger:DIGitize:STIMulus
This command sets the instrument to digitize a measurement the next time it detects the specified trigger event.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
NONE
Usage
[:SENSe[1]]:TRIGger:DIGitize:STIMulus <event>
[:SENSe[1]]:TRIGger:DIGitize:STIMulus?
<event>
The event to use as a stimulus; see
Details
Details
This command is intended to provide the lowest possible latency between a trigger event such as
digital I/O and a reading. It forces the instrument to make a digitize measurement the next time it
detects the specified trigger event. Options for the trigger event parameter are listed in the following
table.
A digitize function must be active before sending this command. The measurement is digitized for the
active function. If a measure function is active, an error is generated.
Before using this command, set the active reading buffer. Readings are stored in the active reading
buffer.
If the count is set to more than 1, the first reading is initialized by this trigger. Subsequent readings
occur as rapidly as the instrument can make them. If a trigger occurs during the group measurement,
the trigger is latched and another group of measurements with the same count will be triggered after
the current group completes.
If the stimulus is set to none, this command has no effect on readings.