Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-48
DMM7510-901-01 Rev. B / May 2015
FORMat subsystem
The commands for this subsystem select the data format that is used to transfer instrument readings
over the remote interface.
:FORMat:ASCii:PRECision
This command sets the precision (number of digits) for all numbers returned in the ASCII format.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0
Usage
:FORMat:ASCii:PRECision <n>
:FORMat:ASCii:PRECision DEFault
:FORMat:ASCii:PRECision MINimum
:FORMat:ASCii:PRECision MAXimum
:FORMat:ASCii:PRECision?
:FORMat:ASCii:PRECision? DEFault
:FORMat:ASCii:PRECision? MINimum
:FORMat:ASCii:PRECision? MAXimum
<n>
The precision:
•
Automatic:
0
•
Specific value:
1
to
16
Details
This attribute specifies the precision (number of digits) for queries.
Note that the precision is the number of significant digits. There is always one digit to the left of the
decimal point; be sure to include this digit when setting the precision.
Example
:FORM:ASC:PREC 10
Set a precision of 10 digits. An example of the
output is:
-6.999999881E-01
Also see