Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-111
[:SENSe[1]]:<function>:THReshold:RANGe:AUTO
This command determines if the threshold range is set manually or automatically.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
ON (1)
Usage
[:SENSe[1]]:<function>:THReshold:RANGe:AUTO <state>
[:SENSe[1]]:<function>:THReshold:RANGe:AUTO?
<function>
The function to which the setting applies; see
Functions
<state>
The auto range setting:
•
Disable:
OFF
or
0
•
Enable:
ON
or
1
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command determines how the range is selected.
When this command is set to off, you must set the range. If you do not set the range, the instrument
remains at the range that was selected by autorange.
When this command is set to on, the instrument uses the signal to determine the most sensitive
range on which to perform the measurement. The instrument sets the range when a measurement is
requested. To set the range, the instrument makes a measurement to determine the range before
making the final measurement, which can result in slower reading times. Turn autorange off and set a
specific range to increase measure time.
If a range is manually selected through the front panel or a remote command, this command is
automatically set to off.
Example
:FUNC "FREQ"
:FREQ:THR:RANG:AUTO OFF
:FREQ:THR:RANG 10
Set measure function to frequency.
Disable the threshold autorange.
Set the range to 10 V.
Also see
[:SENSe[1]]:<function>:THReshold:LEVel
(on page 6-109)
[:SENSe[1]]:<function>:THReshold:RANGe
(on page 6-110)